TRANSIC100LP Gas Transmitter
Brand: Sick AGMeasuring the oxygen in the process can be quite simple. The TRANSIC100LP from SICK is the rugged transmitter that uses Tunable Diode Laser Spectroscopy (TDLS) to measure O2 reliably.
A technology more common in the world of high-tech analysis, is now being used in a field instrument: at an attractive price, easy installation, a long life, low-maintenance, and suitable for use in explosion-hazardous areas.
Features- Measures in real-time directly in the process
- Easy installation and operation
- Self-diagnostics with maintenance display
- Low requirements for gas conditioning
- Low operating costs: no consumables and no purging gas consumption
- Rugged: reliable measurement even in contaminated gases
Specifications
Brand
Certifications & Approvals
- FM
- CE
- NEC/CEC (US/CA)
- IECEx
- ATEX
Measuring Range
- O₂: 0 to 5 Vol. % / 0 to 100 Vol. %
Response Time
- ≤ 10 s
Accuracy
- ≤ 0.2 Vol. %
Process Temperature
- -20° to 80° C
Process Pressure
- 800 to 1,400 hPa
Ambient Temperature
- -20° to 60° C
Stroage Temperature
- -30° to 80° C
Communication Interface
- RS-232 / RS-485
- USB
Materials of Construction
- Polymer Coating
- MgF2
- SiN
- PTFE
- EPDM
- Kalrez
- Stainless Steel 1.4404 (AISI 316L)
Mounting
- For ATEX/IECEx Version via TSA151 Connecting unit, a PELV Power Supply is mandatory
- Wall Mounting Bracket for Ambient Aair Measurement & Measurement with Measuring Gas Cell
- Flange for Direct Installation in the Process
Power Supply
- Current Consumption ≤ 500 mA
- 24 Vdc, TRANSIC151LP 21.6-26.4 Vdc
- Power Consumption ≤ 6 W, TRANSIC151LP: ≤ 5.2 W
- For ATEX/IECEx Version via TSA151 Connecting Unit, a PELV Power Supply is Mandatory
Documentation
Data Sheets
Product Manuals
- Safety information for non-Ex analyzers pdf 2.8 MB
- TRANSIC111LP Operating Instructions pdf 3.2 MB
- TRANSIC111LP Operating Instructions pdf 3.2 MB
- TRANSIC121LP Laser Oxygen Transmitter Operating Instructions pdf 4.5 MB
- TRANSIC151LP, Safety information for Ex Devices pdf 950 KB
- TRANSIC151LP, USB Interface Addendum To Operating Instructions pdf 1.6 MB
- TRANSIC151LP, USB Interface Addendum To Operating Instructions pdf 650 KB