CA450 Process Multimeter
Brand: Yokogawa Electric CorporationThe Yokogawa CA450 Process Multimeter is a portable test instrument to combine a digital RMS multimeter and a loop calibrator to give process technicians the same functionality in one tool.
The Yokogawa CA450 provides 24-Volt loop power, reducing the need for carrying a separate power supply when performing transmitter testing. And with its built-in, selectable 250-Ohm Hart resistor, it also eliminates the need for a separate resistor.
Process technicians can now do the same job while carrying less equipment.
Features- Source commonly used field instrumentation standards such as loop power, and miliamps DC
- Measure like a DMM; capture and log current, voltage, resistance, and even DCS power
- Step/Sweep outputs for enhanced testing of field equipment
- PC communication for traferring and saving data
- Prevent polarity error issues with terminal safeguards to prevent incorrect wiring
- Conforms to 600V CAT. IV and 1000V CAT III
Specifications
Accuracy
- ± (0.05 % of Reading + 2-Digits) - DC Current Measurement
- ± (0.09 % of Reading + 1-Digit) - DC Voltage Measurement
- ± (0.09 % of Reading + 2-Digits) - DC Voltage Measurement
- ± (0.1 % of Reading + 1-Digits) - DC Voltage Measurement
- ± (0.2 % of Reading + 1-Digit) - Resistance Measurement
- ± (0.2 % of Reading + 2-Digits) - Resistance Measurement
- ± (0.35 % of Reading + 3-Digits) - Resistance Measurement
- ± (0.5 % of Reading + 5-Digits), 50/60 Hz - AC Voltage Measurement
- ± (1 % of Reading + 2-Digits) - Resistance Measurement
- ± (1 % of Reading + 5-Digits), 40 to 500 Hz - AC Voltage Measurement
- ± (1.5 % of Reading + 5-Digits), 500 Hz to 1 kHz
- 0.05 % - High-Precision Signal Measurement
Approvals & Certifications
- CE
- EMC Standards: EN 61326-1, EN 61326-2-2 & EN 55011 Class B Group 1
Approvals & Certifications
- EN 61010-2-031
- EN 61010-2-033
Battery Life
- 10 Hours - DC Current Output (SOURCE) 12 mA (500 ohms Load)
- 140 Hours - DC Voltage & DC Current Output (SIMULATE)
Communication
- BRAIN
- HART
Current
- < 1.2 mA - Max.
- 30 to 100 mA - DC Current Measurement
Display
- 5-Digit LCD (7-Segment)
Electrical Rating
- CAT III 1000 V
- CAT IV 600 V
Electromagnetic Rating
- RFI/EMI
Frequency
- 10 Hz to 20,000 kHz
Function
- Auto Hold
- Data Hold
- Maximum
- Measurement - DC Voltage, AC Voltage, DC Current, Resistance, Frequency, Continuity Check, Diode Test
- Peak Holder
Humidity
- 70 % Relative Humidity or Less (Non-Condensing) - Storage
- 80 % Relative Humidity or Less (Non-Condensing) - Operating
Indication
- Low Battery Indicator
- Overrange Indicator: ”OL”
- Polarity Indicator
Input
- 0.3 to 600 Vrms - Frequency Measurement
- 1,000 Vdc - Max.
- 1,000 Vrms - Input Protective Voltage
- 1,000 Vrms AC - Max.
Insulation Resistance
- 100 megaohms or Greater at 1000 Vdc
Load
- 0-20 mA - SOURCE (Compliance Voltage 28 V), SIMULATE (SINK) External Power Supply 15 to 48 V Overrange up to 25 mA < 10 MH
- 24 Vdc Typical, Load Current 20 mA
Operating Temperature
- -20° to 55° C (-4° to 131° F)
Output
- 20 mA - DC Output
- 20 mA DC Current Output Source & Current Output Simulate (Sink)
- 4-20 mA - to Drive the 2-Wire Transmitter
Power Requirement
- 24 V - Loop Power
- 4 AA Alkaline Batteries
Power Supply
- 24 Vdc / 20 mA DC - Loop Power
Range
- 30.000 mA DC - High-precision Signal Measurement
- 600 mV-1,000 V - DC Voltage, AC Voltage Measurement
Resistance
- 250 ohms - HART Mode Resistance
- 600 ohms - Continuity Check
- 600 ohms to 60 megaohms
- Input Resistance: 10 megaohms or more - DC Voltage, AC Voltage (< 200 PF, < 50 PF)
Resolution
- 0.001 mA - DC Current Measurement
- 0.01 mA - DC Current Measurement
- 0.1 mV-1 V - DC Voltage, AC Voltage Measurement
- 0.1 ohms - Continuity Check
- 0.1 ohms to 0.01 megaohms - Resistance Measurement
- 1 Hz - Frequency Measurement
- 1 V - DC & AC Voltage
Response Time
- Within 1 second - DC Voltage, DC Current Measurement
- Within 10 seconds for 6 megaohms to 60 megaohms
- Within 2 seconds for 600 ohms to 600 kiloohms
Shock
- 1 metre Drop
Storage Temperature
- -40° to 70° C (-40° to 158° F)
Supply Voltage
- < 3.5 V - Open-Loop Voltage
- 6.88 kVac for 5 seconds (Between the Input Terminals & the Case) - Withstand Voltage
Temperature Effect
- Add the Accuracy of 23° C ± 5° C x 0.1/°C
Vibration
- 0.15 mm - Sweep Vibration Frequencies 10 Hz to 5 Hz to 10 Hz
Voltage Drop
- < 0.3 V - DC Current Measurement
- < 0.8 V - DC Current
- < 3.5 V - Resistance