DRU Cryogenic Temperature Converter

Brand: Yokogawa Electric Corporation

The DRU is a nest-mounting type DCS-supported cryogenic temperature converter that is used in combination with a four-wire cryogenic RTD (Yokogawa J263*B) and converts high-precision measured values of wide temperature ranges into isolated DC current or DC voltage signals.

Specifications

Accuracy

  • Output 1: ± 0.1 % of Span or ± 0.3° C (± 0.54° F), whichever is greater (Includes Linearization Error)
  • Output 2: within ± 0.2 % Relative Error between Output-1 & 2

Burn-Out

  • Down
  • Off
  • Up

Consumption

  • 70 mA at 24 Vdc - Current Consumption

Electrical Connectors

  • M4 Screw Terminal of the Dedicated Nest - External Wiring

Enclosure, Body Material

  • ABS Resin

Humidity

  • 5 to 90 % Relative Humidity, Non-Condensing

Insulation Resistance

  • 100 megaohms or more at 500 Vdc between Input & Output, Output & Power Supply

Mounting

  • Nest-Mounting

Operating Temperature

  • 0° to 50° C (32° to 122° F)

Output

  • 1-5 Vdc

Power Supply Effect

  • ± 0.1 % of Span or Less for the Fluctuation within the Operating Range of Power Supply Voltage Specification

Resistance

  • 1 Megaohm During Power On - Input Resistance
  • 2 kiloohms or more for Both Output 1 & Output 2 - Load Resistance
  • 3 kiloohms During Power Off - Input Resistance
  • 50 ohms or Less Per Wire - Allowable Leadwire Resistance

Response Time

  • 200 milliseconds, 63 % Response (10 to 90 %)

Sensor, Probe Type

  • 4-Wire RTD

Span & Zero Adjustments

  • ± 1 % - Input Adjustment
  • ± 10 % - Output Adjustment

Supply Voltage

  • 1,500 Vac/minute between Input & Input, (Output & Power Supply) - Withstand Voltage
  • 24 Vdc ± 10 % (Ripple Content 5 % Peak-to-Peak or Less)
  • 500 Vac/minute between Output & Power Supply - Withstand Voltage

Temperature Effect

  • ± 0.2 % of Span or Less or for a Temperature Change of 10° C (18° F) - Effect of Ambient Temperature

show more