SC29 Electrode for pH/Redox
Brand: Yokogawa Electric CorporationThis combined electrode can be used in processes with a constant pH value or for processes where the Redox potential is dependent on the pH in order to achieve a pH compensated Redox potential.
In such a case a pH/Redox converter with a high input is required. All pH/Redox converters of Yokogawa have such an input.
Features- Bulb shaped glass membrane
- Ag/AgCl wire reference system
- Glass resistance: 50 to 100 MegaOhms
- General purpose redox measurements
- Diaphragm resistance (25° C) < 5 kΩ
- High quality Ag/AgCl reference system (pin) which can stand high temperatures and temperature fluctuations
- Double junction, thickened saturated KCl-solution
- Built-in salt bridge prevents poisoning of the reference
- Large area PTFE junction to resist fouling
- Heavy duty pH sensitive glass
- Flowing reference system for pollution resistance, and highly stable reference potential
- PG13.5 standard DIN electrode connection
- Adapter to ensure compatibility with full fitting program
Specifications
Applications
- Chemical: Electroplating
- High Process Pressure Applications
- Pulp & Paper
Approvals & Certifications
- IEC
BSP Connection
- G 1/4” Male
Conduit Connection
- PG 13.5
Diaphragm Material
- Ceramic
- PTFE (Teflon) Membrane
- Universal Glass
Environmental Protection
- IP65
- Watertight
Materials of Construction
- Bulb Shaped Glass Membrane
- Large Area PTFE Junction
- Platinum - Metal Surface
- Silver Chloride Wire - Reference System (Pin)
Maximum System Pressure
- psi: 145 (kPa: 1,000, bar: 10) - SC29-PTG29
- psi: 72.52 (kPa: 500, bar: 5) - SC29C(D)-PTP29, SC29C-PTC55
Measuring Range
- -1,500 to 1,500 mV
NACE & NIST
- NIST
Pressure Range
- psi: 0 to 145 (kPa: 0 to 1,000, bar: 0 to 10)
Resistance
- Diaphragm Resistance [25° C (77° F)]: < 5 kiloohms
- Glass Resistance [25° C (77° F)]: 50 to 100 megaohms
Sensor, Probe Diameter
- 0.47” (12 mm)
Sensor, Probe Length
- 4.72” (120 mm)
Sensor, Probe Material
- Ceramic
- pNa Glass
- PTFE
- Zirconium
Temperature Range
- 0° to 100° C (32° to 212° F) - SC29-PTG29
- -10° to 100° C (14° to 212° F) - SC29C (D)-PTP29, SC29C-PTC55
- Chemical: Electroplating
- High Process Pressure Applications
- Pulp & Paper