Accuracy
- 100 mA Range: 0.015 % of Reading + 30 µA (DC Current Measurement)
- 20 mA & 20 mA Simulate Range: 0.015 % of Setting + 3 µA (DC Current Source)
- 20 mA Range: 0.015 % of Reading + 3 µA (DC Current Measurement)
- 5 V Range: 0.015 % of Reading + 0.5 mV (DC Voltage Measurement)
- 5 V Range: 0.015 % of Setting + 0.5 mV (DC Voltage Source)
- 50 V Range: 0.015 % of Reading + 5 mV (DC Voltage Measurement)
- Negative Pressure: ± (0.2 % of Reading + 0.080 kPa)
- Negative Pressure: ± (0.2 % of Reading + 0.09 kPa)
- Positive Pressure: ± (0.01 % of Reading + 0.003 kPa) for 20 to 200 kPa - CA700-E-01
- Positive Pressure: ± (0.01 % of Reading + 0.010 kPa) for 20 to 200 kPa - CA700-E-01
- Positive Pressure: ± (0.01 % of Reading + 0.04 kPa) - CA700-E-02
- Positive Pressure: ± (0.01 % of Reading + 0.08 kPa) - CA700-E-02
- Positive Pressure: ± (0.01 % of Reading + 0.15 kPa) - CA700-E-03
- Positive Pressure: ± (0.01 % of Reading + 0.30 kPa) - CA700-E-03
- Positive Pressure: ± 0.005 kPa for 0 to 20 kPa - CA700-E-01
- Positive Pressure: ± 0.012 kPa for 0 to 20 kPa - CA700-E-01
Approvals & Certifications
- EMC EN 61326-1 Class A, EN 55011 Class A Group 1
- Safety: EN 61010-1
- Safety: EN 61010-2-030, Contamination Class 2, Measurement Category 0 (Other)
Bandwidth
- 120 dB (50/60 Hz) Approximately - CMRR
- 60 dB (50/60 Hz) - NMRR
Battery Life
- 10 Hours Approximately with the 24 V Loop Power Supply ON
- 35 Hours when Measuring Current with the 24 V Loop Power Supply OFF
BSP Connection
- 1/4” BSPT Female - Input Port
Communication
- USB A Mass-Storage Device
- USB Mini-B Communication Device
Current
- Approximately 36 mA - Generation Unit Current Limiter
Diaphragm Material
- Hastelloy C-276 - Measurement Unit
- SUS316 - Input Port
Display
- Backlight: LED
- Dot Matrix LCD (320 x 240 Dots)
Element
- Diaphragm - Pressure Sensor Element
Environmental Protection
- Dust-Proof
- IP54
- Water-Proof
Function
- Linear - Sweep Function
- Step - Sweep Function
Humidity
- 20 to 80 % Relative Humidity, Non-Condensing - for Operating & Storage Temeprature
Input
- 120 mA - Current Terminal, Max.
- 50 Vdc - Voltage Terminal, Max.
- psi: 0.39 to 435.1 (kPa: 2.7 Absolute to 3,000 Gauge, bar: 0.02 to 30) - Allowable Input for CA700-E-02
- psi: 0.39 to 652.7 (kPa: 2.7 Absolute to 4,500 Gauge, bar: 0.02 to 45) - Allowable Input for CA700-E-03
- psi: 0.39 to 72.52 (kPa: 2.7 Absolute to 500 Gauge, bar: 0.027 to 5) - Allowable Input for CA700-E-01
Insulation Resistance
- 100 megaohms or more at 500 Vdc between the Input Terminal & Case & between the Input Port & Case
Load
- ≤ 0.1 µF - Capacitance
- ≤ 10 MH - Inductance
Maximum System Pressure
- psi: 739.7 (kPa: 5,100, bar: 51)
Measuring Range
- kPa: 240 to 4,200 - Measurement Display Range
Media
- Liquids & Gases (Non-Corrosive, Non-flammable, & Non-Toxic Fluids)
Operating Temperature
- -10° to 50° C (14° to 122° F)
Pressure Range
- Negative Pressure: psig: -11.6 to 0 (kPa: -80 to 0, bar: -0.8 to 0)
- Positive Pressure: psig: 0 to 145 (kPa: 0 to 1,000, bar: 0 to 10) - Model CA700-E-02
- Positive Pressure: psig: 0 to 29.01 (kPa: 0 to 200, bar: 0 to 2) - Model CA700-E-01
- Positive Pressure: psig: 0 to 507.6 (kPa: 0 to 3,500, bar: 0 to 35) - Model CA700-E-03
Resolution
- 0.001 kPa - Model CA700-E-01
- 0.01 kPa - CA700-E-02 & CA700-E-03
- 100 mA Range: 10 µA (DC Current Measurement)
- 20 mA Range & 20 mA Simulate: 1 µA (DC Current Measurement & Current Source)
- 5 V Range: 0.1 mV (DC Voltage Measurement & Voltage Source)
- 50 V Range: 1 mV (DC Voltage Measurement)
Response Time
- 2.5 seconds or Less
- 60 minutes Approximately - Auto Power-Off
Storage Temperature
- -20° to 60° C (-4° to 140° F)
Supply Voltage
- 24 V ± 1 V - Load Current 24 mA when Communication Resistance OFF
- 24 V ± 6 V - Load Current 20 mA when Communication Resistance ON
- 42 V Peak - Generation Unit Voltage to Ground
- 50 V Peak - Measurment Unit Voltage to Ground
- 500 Vac for 1 Minute between the Input Terminal & Case & between the Input Port & Case - Withstand Voltage
- Approximately 36 V - Generation Unit Voltage Limiter
Temperature Effect
- ± (0.001 % of Reading + 0.001 % of Full Scale)/°C or Less - Temperature Coefficient
Temperature Range
- -10° to 50° C (14° to 122° F) - Measurement Fluid
Update Time
- 5 minutes Approximately - Warm-Up
Zero Offset
- ± 0.3 kPa or Less, Zero-Point Drift - Influence of Positional Setup