Accuracy
- AC Voltage: ± (0.5 % of Reading + 5-Digits)
- DC Current: ± (0.025 % + 30 μA) at 23 ± 5° C (73.4 ± 9° F) Per Year
- DC Voltage: ± (0.02 % + 20 mV) at 23 ± 5° C (73.4 ± 9° F) Per Year
- Frequency, Pulse Amplitude Accuracy: ± (5 % + 0.1 V)
- mA SINK: ± (0.05 % + 3 µA) at 23 ± 5° C (73.4 ± 9° F) Per Year
- Resistance: ± (0.025 % + 0.1 ohms) at 23 ± 5° C (73.4 ± 9° F) Per Year
- RTD: ± [0.025 % + 0.3° C (32.54° F)] at 23 ± 5° C (73.4 ± 9° F) Per Year
- Thermocouple: ± [0.02 % + 2.5° C (36.5° F)] 100° C or Less at 23 ± 5° C (73.4 ± 9° F) Per Year
Approvals & Certifications
- EN 55011 Class B Group 1
- EN 61000-3-2
- EN 61000-3-3
- EN 61010-031
- EN 61326-1
Battery Life
- Approximately 12 Hours - Simultaneous Signal Generation / Measurement, Output 20 mA / 5 V
- Approximately 20 Hours - Simultaneous Signal Generation / Measurement, Output 5 Vdc / 10 kiloohms or Greater
- Approximately 40 Hours - Measurement Off, Output 5 Vdc / 10 kiloohms or Greater
Current
- DC Current, Max.
- Excitation Current: 0.5-5 mA
- Frequency, Pulse, Max.
- Signal Generating Unit Current Limiter: Approximately 25 mA
Display
- Backlight: LED - Auto Off after One Minute (From when LIGHT Key is Turned On)
- Segmented LCD (Approximately 76 mm x 48 mm)
Frequency
- AC Voltage Input Frequency: 45 to 65 Hz
- Frequency, Pulse Contact Input: Max.
- Frequency, Pulse: 1 Hz to 11 kHz
Humidity
- 20 to 80 % Relative Humidity, Non-Condensing - for Operating Temperature
- 90 % Relative Humidity or Less (Non-Condensing) - for Storage Temperature
Input
- Current Terminal Input Protection Fuses: 125 mA / 250 V
- Frequency, Pulse, Max.
- Measuring Unit, Max.
Insulation Resistance
- 50 megaohms or Greater, 500 Vdc - Across Input Terminal & Output Terminal
Measuring Range
- Thermocouple: -10° to 50° C (14° to 122° F)
Operating Temperature
- 0° to 50° C (32° to 122° F)
Output
- DC Voltage, Max.
- Frequency, Pulse: 0.1-15 V (Zero Base Waveform)
Power
- Consumed Power: Approximately 7 VA (Using 100 Vac Adapter)
Power Requirement
- 4 AA Alkaline Batteries: A1070EB x 4
Power Supply
- 24 Vdc - Transmitter
- mA SINK External Power Supply: 5-28 V
Range
- AC Voltage: 0-300 V
- DC Current: 0 to ± 100 mA
- DC Current: 0-24 mA
- DC Voltage: 0 to ± 110 V
- DC Voltage: -10 mV to 30 V
- Sink: 0.1-24 mA
Resistance
- AC & DC Voltage Input Resistance: Approximately 10 megaohms/10 PF
- DC Current Input Resistance: Approximately 14 ohms
- Frequency, Pulse Input Resistance: 200 kiloohms or Greater
- Range: 0 to 400 ohms
Resolution
- AC Voltage: 1 V
- DC Current: 4 mA
- DC Voltage: 0.01 V
- Frequency, Pulse: 0.01 Hz to 0.1 kHz
- mA SINK: 1 µA
- Resistance: 0.01 ohms
- RTD: 0.1° C
- Thermocouple: 1° C
Response Time
- Signal Generating Unit Response Time: Approximately 1 second (Time between Start of Voltage Change & when Voltage Enters Accuracy Range)
Sensitivity
- Frequency, Pulse Sensitivity: 0.5 V Peak or Greater
Sensor, Probe Type
- B Thermocouple
- E Thermocouple
- J Thermocouple
- K Thermocouple
- L Thermocouple
- N Thermocouple
- RTD
- R Thermocouple
- S Thermocouple
- T Thermocouple
- U Thermocouple
Storage Temperature
- -20° to 50° C (-4° to 122° F)
Supply Voltage
- Frequency, Pulse, Max.
- Measuring Unit Ground Voltage: Max.
- Signal Generating Unit Current Limiter: Approximately 32 V
- Withstand Voltage: 3,700 Vac for One Minute Across Input Terminal & Output Terminal
Sweep Time
- Sweep Function: Approximately 16 seconds
- Sweep Function: Approximately 32 seconds
Temperature Range
- JPt100: -200° to 500° C (-328° to 932° F)
- Pt100: -200° to 850° C (-328° to 1,562° F)
- Thermocouple: -200° to 1,800° C (-328° to 3,272° F)
Turn-Off Time
- Approximately 10 minutes (Auto-Power-Off Can be Disabled through A DIP Switch Setting)
Update Time
- Measurement Display Updating Rate: Approximately Once Per second