ISC40FDJ Immersion Holder

Brand: Yokogawa Electric Corporation

The ISC40J sensors are designed for use with the FLXA202/FLXA21 2-wire Analyzer, ISC202 2-wire transmitter and the ISC450G 4-wire converter.

This combination exceeds all expectations for conductivity measurement in terms of reliability, accuracy, rangeability and price performance.

This innovative inductive conductivity sensor provides highly accurate measurements over a wide measuring range (1 µS/cm to 1999 mS/cm) and process temperature range (-10° to 130° C, -10° to 105° C for ISC40SJ-TT) without changing the cell constant and conducting re-calibration.
Features
  • Inductive Conductivity technique for elimination of fouling and polarization errors.
  • Wide bore sensors for long term stability.
  • Installation flexibility due to wide range of holders and due to the use of universal bulkhead construction.
  • A single sensor can maintain the high resolution and accuracy, and measure the conductivity in an extremely brood range.
    • Minimum span: 100 μS/cm
    • Maximum span: 1999 mS/cm

Specifications

Applications

  • All Applications where Severe Electrode Fouling Prevents the Use of Contacting Electrodes
  • All Applications where the 6 Decade Rangeability is Necessary for Accurate Process Control
  • All Ranges Except (Ultra) Pure Water Applications
  • All Slurry Applications where Conventional Systems Suffer from Plugging or Erosion

Enclosure, Body Material

  • 316 Stainless Steel
  • C-PVC

Environmental Protection

  • Watertight

Flange Material

  • 316 Stainless Steel
  • PP

Flange Size - 2” (DN 50)

  • 2” (DN 50) - 150# ANSI
  • 2” (DN 50) - JIS 10K
  • 2” (DN 50) - PN 10

Gasket Material

  • Chloroprene
  • Ethylene Propylene Copolymer Rubber

Materials of Construction

  • Zinc-Plated Steel - Mounting Set

Maximum System Pressure

  • psi: 14.5 (kPa: 100, bar: 1) at 80° C
  • psi: 29 (kPa: 200, bar: 2) at 20° C

Mounting

  • 2” Pipe Mounting

O-Ring Material

  • Ethylene Propylene Copolymer Rubber
  • Fluororubber

Process Temperature

  • 80° C (176° F) - Max.

Certifications & Approvals

  • TIIS

  • All Applications Where Severe Electrode Fouling Prevents the Use of Contacting Electrodes
  • All Applications Where the 6 Decade Rangeability is Necessary for Accurate Process Control
  • All Ranges Except (Ultra) Pure Water Applications
  • All Slurry Applications Where Conventional Systems Suffer from Plugging or Erosion
show more