OR8EFG KCI Filling Type ORP Sensor

Brand: Yokogawa Electric Corporation

Yokogawa’s process OR8EFG meters are highly reliable and feature advanced functions which are useful for a wide variety of applications including water quality management in a broad range of production processes and medium-sized wastewater treatment plants, or for general pH and ORP control systems.

Based on Yokogawa’s track record and years of experience, a comprehensive range of products has been produced to provide solutions best suited to individual applications.
Features
  • With the body made of Ryton, a strong engineering plastic, which is comparable to Fluoropolymer (PTFE) in terms of corrosion resistance and heat resistance, it allows for a wide range of applications
  • A single type of sensor can support all applications regardless of whether a holder or cleaner is used
  • The integrated-sensor design simplifies calibration with standard solutions and maintenance
  • The platinum or gold electrode of a ORP sensor and junction can be individually replaced

Specifications

Applications

  • General ORP Control Systems
  • Medium-Sized Waste Water Treatment Plants
  • Water Quality Management in a Broad Range of Production Processes

Cable Length

  • 9.84 to 65.61 feet (3 to 20 metres)

Cable Material

  • Chlorinated Polyethylene Rubber - Cable Sheath

Electrical Connectors

  • G 1/2” Plastic Gland: 21 mm (pH Sensor Side) & 13 mm (pH Transmitter or Converter Side) Terminal Box

Enclosure, Body Material

  • Ceramic
  • Fluoro Rubber (FKM)
  • Gold-Epoxy Resin
  • Hastelloy C
  • Platinum-glass
  • Ryton (PPS Resin)
  • Titanium

Materials of Construction

  • Gold - Electrode
  • Platinum - Electrode
  • Polyethylene (Medium Pressure) - KCl Tube
  • Polypropylene - Holder Material for Float, Submersion, Guide-Pipe
  • PVC - Holder Material for Guide-pipe & KCl Tube
  • Stainless Steel - Holder Material for Float, Suspension, Flow-through, Submersion

Measuring Range

  • -1,500 to 1,500 mV

Mounting

  • 2” Pipe Mounting Bracket

Pressure Range

  • psi: 0 to 1.45 (kPa: 0 to 10, bar: 0 to 0.1) - General-purpose Atmospheric Pressure
  • psi: 0 to 72.52 (kPa: 0 to 500, bar: 0 to 5) - Medium Pressure Atmospheric Pressure

Principle of Operation

  • Metal Electrode Method

Process Temperature

  • -5° to 105° C (23° to 221° F)

Sensor, Probe Style

  • Float

  • General ORP Control Systems
  • Medium-Sized Waste Water Treatment Plants
  • Water Quality Management in a Broad Range of Production Processes
show more