SC29 Electrode for pH/Redox

Brand: Yokogawa Electric Corporation

This combined electrode can be used in processes with a constant pH value or for processes where the Redox potential is dependent on the pH in order to achieve a pH compensated Redox potential.

In such a case a pH/Redox converter with a high input is required. All pH/Redox converters of Yokogawa have such an input.
Features
  • Bulb shaped glass membrane
  • Ag/AgCl wire reference system
  • Glass resistance: 50 to 100 MegaOhms
  • General purpose redox measurements
  • Diaphragm resistance (25° C) < 5 kΩ
  • High quality Ag/AgCl reference system (pin) which can stand high temperatures and temperature fluctuations
  • Double junction, thickened saturated KCl-solution
  • Built-in salt bridge prevents poisoning of the reference
  • Large area PTFE junction to resist fouling
  • Heavy duty pH sensitive glass
  • Flowing reference system for pollution resistance, and highly stable reference potential
  • PG13.5 standard DIN electrode connection
  • Adapter to ensure compatibility with full fitting program

Specifications

Applications

  • Chemical: Electroplating
  • High Process Pressure Applications
  • Pulp & Paper

Approvals & Certifications

  • IEC

BSP Connection

  • G 1/4” Male

Conduit Connection

  • PG 13.5

Diaphragm Material

  • Ceramic
  • PTFE (Teflon) Membrane
  • Universal Glass

Environmental Protection

  • IP65
  • Watertight

Materials of Construction

  • Bulb Shaped Glass Membrane
  • Large Area PTFE Junction
  • Platinum - Metal Surface
  • Silver Chloride Wire - Reference System (Pin)

Maximum System Pressure

  • psi: 145 (kPa: 1,000, bar: 10) - SC29-PTG29
  • psi: 72.52 (kPa: 500, bar: 5) - SC29C(D)-PTP29, SC29C-PTC55

Measuring Range

  • -1,500 to 1,500 mV

NACE & NIST

  • NIST

Pressure Range

  • psi: 0 to 145 (kPa: 0 to 1,000, bar: 0 to 10)

Resistance

  • Diaphragm Resistance [25° C (77° F)]: < 5 kiloohms
  • Glass Resistance [25° C (77° F)]: 50 to 100 megaohms

Sensor, Probe Diameter

  • 0.47” (12 mm)

Sensor, Probe Length

  • 4.72” (120 mm)

Sensor, Probe Material

  • Ceramic
  • pNa Glass
  • PTFE
  • Zirconium

Temperature Range

  • 0° to 100° C (32° to 212° F) - SC29-PTG29
  • -10° to 100° C (14° to 212° F) - SC29C (D)-PTP29, SC29C-PTC55

  • Chemical: Electroplating
  • High Process Pressure Applications
  • Pulp & Paper
show more