IR200 NDIR Type Infrared Gas Analyzer
Brand: Yokogawa Electric CorporationThe IR200 infrared gas analyzer is capable of measuring the concentrations of CO₂, CO, CH₄, SO₂, NO and O₂ components in sample gas.
CO₂, CO, CH₄, SO₂ and NO are measured by nondispersive infrared method (NDIR), while O₂ is measured by paramagnetic or zirconia method.
A maximum of 4 components including O₂ (up to 3 components except for O₂ measurement) are simultaneously measurable.
A high-sensitivity mass flow sensor is adopted in the detection unit for the infrared method.
Due to use of single beam system for measurement, maintenance is easy and an excellent stability is ensured for a long period of time.
In addition, the IR200 and has a large-size liquid crystal display, providing easy operation, high accuracy and multiple functions.
Features- Simultaneous measurement of 4 components including O2
- Excellent long-term stability
- Subdued interference from other gas components
- Easy to maintain
- Easy to operate
- Abundant functions
Specifications
Ambient Temperature Range
- -20° to 60° C (-4° to 140° F) - for Storage Conditions
- -5° to 45° C (23° to 113° F) - for Operating Conditions
Approvals & Certifications
- EMC - EN 61326 Class A, EN 61326-2-3, EN 61000-3-2, EN 61000-3-3
- RCM - EN 61326-1 Class A
BSP Inlet
- 1/4" BSPT Female
BSP Outlet
- 1/4" BSPT Female
Cable Length
- 8.2 ft (2.5 m) - Power Cable
Communication
- Half-Duplex Bit Serial (Start-Stop Synchronization)
- Modbus Protocol
- RS-232C (9-Pins D-sub)
- RS-485
Consumption
- 70 VA, Max.
Display
- 4-Digit LCD with CFL Backlight
Electrical Connectors
- M3.5 Screw Terminals
Enclosure Finish
- Off-White (Munsell 10Y7.5/0.5 or Equivalent) - Front Panel
Enclosure, Body Material
- Steel - Blue, for Indoor Use
Flow Rate
- 0.26 gpm (1 LPM) - for Purge Gas
- 1 ± 0.5 LPM - for Sample Gas
Frequency
- 50/60 Hz
Humidity
- 90 % RH, Non-Condensing, max.
Input Signal
- 0-1 Vdc from an O₂ Sensor Input Section is Non-Isolated - Analog Input Signal
Interface Connection
- RS-485
Linearity
- ± 1 % of Full Scale
Load
- 100 kΩ min. for 0-1 Vdc - Permissible Load Resistance
- 550 ohms, Max.
Materials of Construction
- 304 Stainless Steel - Gas Inlet/Outlet, Sample Cell
- Neoprene Rubber - Sample Cell
- Toalon Tube - Internal Tubing
Measuring Range
- 0 to 1,000 ppm Min., 0 to 100 Vol %, Max.
- 0 to 200 ppm Min., 0 to 100 Vol %, Max.
- 0 to 5 Vol % Min., 0 to 100 Vol %, Max.
- 0 to 5 Vol % Min., 0 to 25 Vol %, Max.
- 0 to 500 ppm Min., 0 to 100 Vol %, Max.
- 0 to 500 ppm Min., 0 to 5,000 ppm, Max.
- Dust: 100 µg/Nm³ or Less in Particle Size of 0.3 μm or Less
Media
- Dry Air
- Gas
Mounting
- 19" Rack Mounting with Slide Rail
NPT Inlet
- 1/4" NPT Female - Gas Inlet
NPT Outlet
- 1/4" NPT Female - Gas Outlet
Operating Temperature
- -5° to 45° C (23° to 113° F)
Output
- 0-1 Vdc, Non-Isolated - Analog Output Signal
- 4-20 mA DC, Non-Isolated - Analog Output Signal
Power Supply
- 100-240 Vac - Voltage Rating
- 85-264 Vac - Allowable Range
Pressure Range
- psi: 0 to 1.45 (kPa: 0 to 10, bar: 0 to 0.1) or Less (Gas Outlet Side Should be Open to the Atmospheric Air)
Ratio
- ≤ 1:20 (Built-in Paramagnetic O₂ Analyzer) - Measuring Range Ratio
- ≤ 1:5 (Except Built-in Paramagnetic O₂ Analyzer) - Measuring Range Ratio
Relay Output
- 250 Vac / 2 A, Resistive Load - Relay Contact Output
Repeatability
- ± 0.5 % of Full Scale
Response Time
- Within 15 seconds Including Replacement Time of Sample Gas (For 90 % of Full Scale Response) - for 1 or 2 Component Measurement
- Within 30 seconds Including Replacement Time of Sample Gas (For 90 % of Full Scale Response) - for more than 2 Component Measurement
Span & Zero Effect
- ± 2 % of Full Scale Per Week
Temperature Range
- 0° to 50° C (32° to 122° F)
Window Material
- CaF₂ - Infrared Ray Transmitting Window