DM7560 Digital Multimeter
Brand: Yokogawa Electric CorporationThe DM7560 is a flexible 6.5-digit digital multimeter with high-accuracy, high-speed data logging (up to 30 kS/s) with deep 100 k reading memory and built-in statistical analysis.
Features- High resolution up to 6-1/2 digits
- Full-color, high-resolution display with flexible display formats
- High-speed data logging with up to 30 k points per second
- High-capacity internal memory up to 100 k points
- Trend/histogram analysis available both in real-time and offline
- Multiple PC interface options (USB, Ethernet, GP-IB, RS-232) enable automation
Specifications
Accuracy
- ± (0.0050 % of Reading + 0.0035 % of Range) - DC Voltage
- ± (0.010 % of Reading + 0.020 % of Range) - Continuity & Diode Test
- ± (0.120 % of Reading + 0.020 % of Range) - DC Current
- ± (0.20 % of Reading + 0.70° C (1.26° F)) - Thermocouple
- ± (0.70 % of Reading + 0.04 % of Range) - AC Current
- ± (0.800 % of Reading + 0.010 % of Range) - Terminal Resistance
- ± (4 % of Reading + 0.50 % of Range) - AC Voltage
- ± 0.06° C (± 0.1° F) - RTD
- 0.1 % of Reading - Frequency
Applications
- Monitoring Battery Current Consumption
- Production Test
- Research Institutions R&D/Service
- Testing Sensor
- Voltage Reference Testing
Approvals & Certifications
- AS
- BS
- CSA
- GB
- NBR
- PSE
- UL
- VDE
Communication
- Bit Rate: 300, 600, 1200, 2400, 4800, 9600, 19200, 38400 bit/sec
- LAN Setting: DHCP (ON/OFF), IP Address, Gateway, Subnet Mask
- RS-232
- Stop Bit: 1-bit, 2-bits
- USB
Consumption
- 21 VA or Less - Power Consumption
Current
- 500 NA -1 mA - Terminal Resistance
Delay Time
- < 1 μs - Trigger Input (BNC)
- 0 to 3,600 seconds Setting Range - Trigger Delay & Interval
Display
- 4.3” (109 mm): TFT Active Matrix
- 5-1/2” Digits
- 6-1/2” Digits
- Backlight: LED
- LCD Display
- Sampling Method: 2.5 to 30 K S/s (DC Function), 2.5 to 5 S/s (AC Filter), 100 S/s (Continuity & Diode Test)
Frequency
- 20 Hz to 300 kHz - AC Voltage
- 20 Hz to 5 kHz - AC Current
- 3 Hz to 300 kHz (Gate Time: 1 millisecond to 1 second)
- 50/60 Hz - Power Frequency
Humidity
- 80 % Relative Humidity (No Dew Allowed Below the Moisture Amount) - Calculation Guaranteed Temperature
- 90 % Relative Humidity (No Dew Allowed Below the Moisture Amount) - Storage Temperature
Input
- 0.6 V Max.
- 0.9 V Max.
- 1 mA-3 A - DC Current
- 100 mV-1,000 V - DC Voltage
- 100 mV-750 V - AC Voltage
- 1-3 mA - AC Current
- 2.4 V Min. - Trigger Input (BNC), Trigger Inhibit Input (DIO Option)
Input Impedance
- 0.1 ohms - AC Current
- 0.1 to 90 ohms - DC Current
- 1 GigaOhms or more or 10 megaohms ± 1 % for 0.1-10 V Range - DC Voltage
- 1 Megaohm / 100 PF or Less Approximately - AC Voltage
- 10 kiloohms - Trigger Input (BNC)
- 10 megaohms ± 1 % for 100-1,000 V Range - DC Voltage
- 5 kiloohms - Trigger Inhibit Input (DIO Option)
Input Signal
- AC & DC Current
- AC & DC Voltage
Interface Connection
- DIO Interface
- GPIB
Maximum Current
- 100 mA - Max.
- 3 A (Continuous, Protection by 3 A Fuse) - Max.
Maximum Voltage
- 1,100 Vpeak (1 Minute) - Max.
- 750 Vrms or ± 500 V or Less, Max.
- 800 Vpeak (Continuous) - Max.
Output
- 0.4 Vmaximum - Complete Output (BNC)
- 2.4 Vminimum - Complete Output (BNC)
Output Impedence
- 1 Kiloohm - COMPLETE Output (BNC)
Output Noise
- 120 Db at 50/60 Hz, ± 0.1 % for CMRR - Noise Rejection in Multiple of 1 PLC
- 55 dB at 50/60 Hz, ± 0.1 % for NMRR - Noise Rejection in Multiple of 1 PLC
Power Supply
- 100 Vac
- 115 Vac
- 220 Vac
- 240 Vac ± 10 %
- 50/60 Hz
Pulse Range
- 1 microsecond or more - Trigger Input (BNC)
- 10 µs at Limit Judgment Off - Output Pulse Width (BNC)
- 4 milliseconds or more at Limit Judgment ON - Output Pulse Width (BNC)
Resistance
- 1 Kiloohm - Continuity Test
- 100 ohms to 100 megaohms - 2 (2 W ohms) & 4 (4 W ohms) Terminal Resistance
Resolution
- 0.001° C - Thermocouple
- 0.01° C - RTD
- 1 mV - AC & DC Voltage
- 1 % - Bulk Mode
- 10 µs - Trigger Function
- 10 milliohms - Continuity Test
- 10 μA - AC & DC Current
- 100 ohms - Terminal Resistance
Response Time
- Within 1 second - DC Voltage
- Within 2 seconds - AC Filter (HIGH)
- Within 3 seconds - AC Filter (MID)
RTD Type
- JPt100
- Pt100
Sensor, Probe Type
- E Thermocouple (CRC)
- J Thermocouple (IC)
- K Thermocouple (CA)
- RTD
- R Thermocouple
- T Thermocouple (CC)
Storage Temperature
- -20° to 60° C (-4° to 140° F)
Supply Voltage
- ± 500 Vdc - Withstand Voltage (Between LO Terminal & Ground)
- 42 Vpeak - Limit Judge Output (DIO Option)
Temperature Coefficient
- ± (0.001 % of Reading + 0.002 % of Range)/°C - Continuity & Diode Test
- ± (0.0050 % of Reading + 0.0005 % of Range)/°C - DC Voltage
- ± (0.100 % of Reading + 0.006 % of Range)/°C - AC Current
- ± (0.200 % of Reading + 0.020 % of Range)/°C - AC Voltage
- ± 0.003° C/°C - RTD
Temperature Range
- 0° to 50° C (32° to 122° F) - Calculation Guaranteed Temperature
- -200° to 1,765° C (-328° to 3,209° F) - Thermocouple
- -200° to 850° C (-328° to 1,562° F) - RTD
Update Time
- 1-Hour after Power On - Warm-Up Time