FLXA21 PROFIBUS PA 2-Wire Analyzer
Brand: Yokogawa Electric CorporationThe FLXA21 is a next-generation modular liquid analyzer that can be flexibly configured to measure pH/ORP, contacting conductivity, inductive conductivity, or dissolved oxygen.
The FLXA21 also supports the installation of up to two sensors of the same type, thereby reducing installation costs and saving space in addition to enabling the configuration of a highly reliable redundant system.
It has reliable and advanced features and functions such as a touch screen for improved operability, sensor self-diagnosis, maintenance time estimation, and 12 language display options.
Features- 4 kinds of measurements; pH/ORP, SC, ISC and DO
- Dual sensor measurement on 2-wire type analyzer; pH/ORP and pH/ORP, SC and SC, and DO and DO
- Calculated data from dual sensor measurement
- Redundant system on dual sensor measurement
- Connection of digital FU20F pH/ORP SENCOM Sensor
- Easy touch screen operation on 2-wire type analyzer
- Simple HMI menu structure in 12 languages
- Quick setup menu for immediate measurement
- Indication of sensor wellness
- Hazardous location approvals – ATEX, IECEx, FM, CSA, NEPSI and KOSHA
Specifications
Accuracy
- ± 0.3° C (± 0.54° F) - Temperature with Pt1000, 6k8, PTC 10k, NTC 8k55, 3k Balco, PTC500, Pb36, Ni100 for PH/ORP with Analog Sensors, Conductivity
- ± 0.4° C (± 0.72° F) - Temperature with Pt100, NTC 8k55 for PH/ORP with Analog Sensors, Conductivity
- ± 0.5 % of Reading - Conductivity, Resistivity
- ± 1 % of Reading - Conductivity, Resistivity
- ORP: ± 1 mV - for PH/ORP with Analog Sensor
- pH: ± 0.01 PH - for PH/ORP with Analog Sensor
Ambient Temperature Range
- -20° to 55° C (-5° to 130° F) - Operating
Approvals & Certifications
- ATEX: Intrinsically Safe
- CE
- CSA: Intrinsic Safety & Non-Incendive
- EMC: EN 61326-1 Class A,Table 2 (For use in Industrial Locations), EN 61326-2-3, EN 61326-2-5, AS/NZS CISPR11, RCM: EN 55011 Class A, Group 1, Korea Electromagnetic Conformity Standard Class A
- FM: Intrinsic Safety & Non-Incendive
- IEC Ex: Intrinsic Safety
- NAMUR NE107
- NEPSI: Intrinsic Safety
- UL 61010-1, 61010-2-030
Cable Entry
- 3 Holes, M20 Cable Gland x 3-Pieces
Cable Length
- 200 feet (60 metres) - Conductivity
Communication
- HART
- PROFIBUS PA
- RS-485 - Bi-Directional Digital Communication between FU20F & FLXA201
Conduit Connection
- 1/2” NPT Female - Conduit Adapter
- G 1/2” Female - Conduit Adapter
- M20 x 1.5 Female - Conduit Adapter
Current
- Current Draw: 24 mA - Max.
Diameter
- 0.07” (1.9 mm) - Max.
- 0.13” to 0.27” (3.4 to 7 mm) - Outer Diameter for Grounding Cable
- 0.23” to 0.47” (6 to 12 mm) - Outer Diameter for M20 Cable Gland
Display
- LCD with a Touch Screen (Black/White: 213 x 160 Pixels)
Electrical Connectors
- M4 Screw Terminal
- Pin Terminal with Diameter 0.07” (1.9 mm) - Max.
- Ring Terminal, Width 0.30” (7.8 mm) - Max.
- Spade Terminal, Width 0.30” (7.8 mm) - Max.
Enclosure Finish
- Silver Gray (Equivalent to Munsell 3.2PB7.4/1.2) - Case Color
- Stainless Steel with Epoxy Coating - Hood
- Stainless Steel with Urethane Coating - Hood
Enclosure, Body Material
- Plastic (Polycarbonate)
Environmental Protection
- Hazardous Area
- Intrinsically Safe
- IP66 (Except Canada & USA)
- NEMA 3S (Canada, USA)
- NEMA 4X (Canada, USA)
- Non-Incendive
Humidity
- 0 to 100 % Non-Condensing Relative Humidity - Ambient
- 10 to 90 % Non-Condensing Relative Humidity at 40° C (104° F)
Input
- ORP: -1,500 to 1,500 mV - for PH/ORP with Analog Sensors & Digital Sensor
- pH: 0 to 14 PH - for PH/ORP with Digital Sensor
- pH: -2 to 16 PH - for PH/ORP with Analog Sensors
- PTC/NTC Thermistor
- rH: 0 to 100 RH - for PH/ORP with Analog Sensor & Digital Sensor
Input Impedance
- Dual High Input Impedance ≥ 10¹² ohms - for PH/ORP with Analog Sensors
Linearity
- ± 0.01 PH
- ± 0.4° C (± 0.72° F)
- ± 1 mV
Materials of Construction
- Hood & Tag Plate: Stainless Steel
Mounting
- Panel Mounting Hardware
- Pipe & Wall Mounting Hardware
- Universal Mounting Kit
Pressure Range
- psi: 11.6 to 15.95 (kPa: 80 to 110, bar: 0.8 to 1.1) - Atmosphere Pressure
Repeatability
- ± 0.1° C (± 0.18° F) - Temperature with Pt1000, 6k8, PTC10k, NTC 8k55, 3k Balco, PTC500, Pt100
- ORP: ± 1 mV
- pH: ± 0.01 PH
Response Time
- 90 % (< 2 Decades) in 7 seconds (of Reading on the Display) - Step Response for Conductivity
RTD Type
- Ni100
- Pt100
- Pt1000
Sensor, Probe Type
- NTC 8k55
- NTC30k
- Pb36 (JIS NTC 6k)
- PTC10k
- PTC500
Storage Temperature
- -30° to 70° C (-22° to 158° F)
Supply Voltage
- 9-32 Vdc
Temperature Range
- -10° to 105° C (14° to 221° F) - PH/ORP with Digital Sensor, FU20F PH/ORP SENCOM Sensor
- -20° to 250° C (-4° to 482° F) - Conductivity
- -30° to 140° C (-22° to 284° F) - PH/ORP with Analog Sensors
Torque
- 10.62 In/Lbs (1.2 Nm) - Torque of Screw Up
Window Material
- Polycarbonate (Flexible)