EJX630A High Performance Pressure Transmitter

Brand: Yokogawa Electric Corporation

The high performance gauge pressure transmitter EJX630A and absolute pressure transmitter EJX610A feature single crystal silicon resonant sensor and are suitable to measure liquid, gas, or steam pressure.

EJX630A output a 4-20 mA DC signal corresponding to the measured pressure. They also feature quick response, remote setup and monitoring via BRAIN or HART communications, diagnostics, and optional status output for pressure high/low alarm.

The multi-sensing technology provides the advanced diagnostic function to detect such abnormalities as an impulse line blockage or heat trace breakage. 

FOUNDATION Fieldbus and PROFIBUS PA protocol types are also available. All EJX series models in their standard configuration, with the exception of the Fieldbus and PROFIBUS types, are certified by TÜV as complying with SIL 2 for safety requirement.
Features
  • Accuracy: ± 0.04% Accuracy (Optional ± 0.025%)
  • Response Time: 90 msec
  • Stability: ±0.1% of URL per 15 years

Specifications

Accuracy

  • ± (0.005 + 0.002 URL/Span) % of Span (Includes the Effects of Terminal-based Linearity, Hysteresis, & Repeatability) - Capsule B, C & D (High Accuracy)
  • ± (0.005 + 0.0035 URL/Span) % of Span (Includes the Effects of Terminal-based Linearity, Hysteresis, & Repeatability) - Capsule B, C & D
  • ± (0.008 + 0.006 URL/Span) % of Span (Includes the Effects of Terminal-based Linearity, Hysteresis, & Repeatability) - Capsule A (High Accuracy)
  • ± (0.02 + 0.007 URL/Span) % of Span (Includes the Effects of Terminal-based Linearity, Hysteresis, & Repeatability) - Capsule A
  • ± 0.025 % of Span (Includes the Effects of Terminal-based Linearity, Hysteresis, & Repeatability) - High Accuracy
  • ± 0.04 % of Span (Includes the Effects of Terminal-based Linearity, Hysteresis, & Repeatability)

Adjustment Type

  • Push Button
  • Zero Adjustment Screw

Alarm

  • Down-Scale: -5 %, 3.2 mA DC or Less
  • Up-Scale: 110 %, 21.6 mA DC or more

Ambient Temperature Range

  • -30° to 80° C (-22° to 176° F) with LCD Display
  • -40° to 85° C (-40° to 185° F)

Approvals & Certifications

  • Leak Test
  • Material Certificate
  • Pressure Test
  • SIL 2 Capability for Single Transmitter Use, SIL 3 Capability for Dual Transmitter Use, IEC 61508: 2000
  • TÜV (Technischer Uberwachungsverein)

Approvals & Certifications

  • ATEX - Flameproof & Intrinsically Safe
  • CE 0038 - Category III, Module H
  • CSA - Explosion-proof & Intrinsically Safe
  • EMC - EN 61326-1 Class A (Industrial Locations), EN 61326-2-3 & EN 61326-2-5 (Fieldbus)
  • FM - Explosion-proof & Intrinsically Safe
  • IECEx - Flameproof Approval & Intrinsically Safe
  • NAMUR NE43
  • PED 2014/68/EU, 97/23/EC

BSP Connection

  • G 1/2” Male DIN 16 288

Burst Pressure

  • psi: 26,400 (kPa: 182,000, bar: 1,820) - D Capsule
  • psi: 7,252 (kPa: 50,000, bar: 500) - A, B & C Capsule

Cable Material

  • Polyethylene Insulated PVC Sheathed

Communication

  • BRAIN
  • FOUNDATION Fieldbus
  • HART FSK Protocol
  • PROFIBUS PA

Connection Material

  • 316L Stainless Steel
  • Hastelloy C-276

Current

  • 6,000 A (1 x 40 μs) - Allowable Current
  • Repeating 1,000 A (1 x 40 μs) 100 times - Allowable Current

Damping Adjustment

  • 0 to 100 seconds

Diagnostics

  • Configuration Error
  • CPU Failure
  • Hardware Failure
  • Heat Trace Monitoring
  • Impulse Line Blockage Detection
  • process Alarm for Pressure or Capsule Temperature
  • User-Configurable Process High/Low Alarm for Pressure

Diaphragm Material

  • Hastelloy C-276

Display

  • 5-Digit Numerical Display
  • 6-Digit Unit Display
  • Bar Graph
  • LCD Display
  • User-Configurable 10-Segment Signal Characterizer (4-20 mA Output)

Distance

  • Up to 2 Kilometers - Communication Distance

Electrical Connectors

  • 1/2” NPT Female
  • G 1/2” Female
  • M20 Female

Electrical Protection

  • Lightning Protection

Enclosure Finish

  • Anti-Corrosion Coating
  • Polyurethane Paint, Mint Green Paint (Munsell 5.6BG 3.3/2.9 or its Equivalent)

Enclosure, Body Material

  • ASTM CF-8M Stainless Steel
  • Low Copper Cast Aluminum Alloy

Environmental Protection

  • IP66
  • IP67
  • NEMA 4X

Fill Material

  • Fluorinated Oil
  • Silicone

Humidity

  • 0 to 100 % Relative Humidity - Ambient Humidity Limit

Input Impedance

  • 10 kiloohms or more at 2.4 kHz - Communicating Device

Load

  • 0 to 1,290 ohms - for Operation
  • 0.22 μF or Less - Load Capacitance
  • 250 to 600 ohms - for Digital Communication
  • 3.3 MH or Less - Load Inductance

Materials of Construction

  • 316 Stainless Steel - Name Plate & Tag, Mounting Bracket
  • Polypropylene - Pipe

Media

  • Gas
  • Liquids
  • Steam

Metric Connection

  • M20 x 1.5 Male DIN 16288

Mounting

  • 2” (50.8 mm) Pipe

Mounting Position Effect

  • Tilting up to 90° will cause Zero Shift up to 0.21 kPa

NACE & NIST

  • NACE MR-0103
  • NACE MR-0175, ISO 15156

NPT Connection

  • 1/2” NPT Female
  • 1/2” NPT Male

O-Ring Material

  • Buna-N - Cover O-Ring
  • Fluororubber - Cover O-Ring

Output

  • 3.6-21.6 mA - Output Range
  • 4-20 mA DC (2-Wire) - Output Signal

Output Format

  • One Transistor Contact Output at Rating: 10.5-30 Vdc, 120 mA DC, Max.

Power Requirement

  • 24 Vdc Supply, up to A 550 ohms Load

Power Supply Effect

  • ± 0.005 % Per Volt (21.6-32 Vdc, 350 ohms)

Pressure Range

  • bar: -1 to 700
  • kPa: -100 to 70,000
  • psi: -14.5 to 10,150

Pressure Type

  • Gauge

Process Connection Material

  • 316L Stainless Steel
  • Hastelloy C-276

Process Temperature

  • -40° to 120° C (-40° to 248° F)

Response Time

  • 90 milliseconds

Screw Material

  • 316 Stainless Steel - Zero-Adjustment Screw & Set Screws

Stability

  • ± 0.1 % of URL Per 10-Years

Supply Pressure

  • psi: 0.38 to 14.5 (kPa: 2.7 to 100, bar: 0.026 to 1)

Supply Voltage

  • 10.5-30 Vdc - for Intrinsically Safe, Typical N, Non-Incendive or Non-Sparking Typical
  • 10.5-32 Vdc - for Lightning Protector
  • 10.5-42 Vdc - for General Use & Flameproof Typical
  • 16.6 Vdc Min. - Digital Communications, BRAIN & HART

Temperature Effect

  • ± (0.04 % of Span + 0.009 % of URL) Per 28° C (50° F) - Ambient Temperature Effect for C & D Capsule
  • ± (0.04 % of Span + 0.018 % of URL) Per 28° C (50° F) - Ambient Temperature Effect for B Capsule
  • ± (0.04 % of Span + 0.075 % of URL) Per 28° C (50° F) - Ambient Temperature Effect for a Capsule

Update Time

  • 45 milliseconds

Vibration Effect

  • < ± 0.1 % of URL as Per IEC 60770-1 Pipeline with Low Vibration Level (10-60 Hz, 0.15 mm Displacement / 60-500 Hz 2 G)
  • < 0.1 % of URL as Per IEC 60770-1 Pipeline with High Vibration Level (10-60 Hz, 0.21 mm Displacement / 60 to 2,000 Hz 3 G)

Wetted Materials

  • 316L Stainless Steel
  • Hastelloy C-276

Zero Offset

  • 0.01 % Incremental Resolution of Span - External Zero Adjustment
show more