IR200 NDIR Type Infrared Gas Analyzer

Brand: Yokogawa Electric Corporation

The IR200 infrared gas analyzer is capable of measuring the concentrations of CO₂, CO, CH₄, SO₂, NO and O₂ components in sample gas.

CO₂, CO, CH₄, SO₂ and NO are measured by nondispersive infrared method (NDIR), while O₂ is measured by paramagnetic or zirconia method.

A maximum of 4 components including O₂ (up to 3 components except for O₂ measurement) are simultaneously measurable.

A high-sensitivity mass flow sensor is adopted in the detection unit for the infrared method.

Due to use of single beam system for measurement, maintenance is easy and an excellent stability is ensured for a long period of time.

In addition, the IR200 and has a large-size liquid crystal display, providing easy operation, high accuracy and multiple functions.
Features
  • Simultaneous measurement of 4 components including O2
  • Excellent long-term stability
  • Subdued interference from other gas components
  • Easy to maintain
  • Easy to operate
  • Abundant functions

Specifications

Ambient Temperature Range

  • -20° to 60° C (-4° to 140° F) - for Storage Conditions
  • -5° to 45° C (23° to 113° F) - for Operating Conditions

Approvals & Certifications

  • EMC - EN 61326 Class A, EN 61326-2-3, EN 61000-3-2, EN 61000-3-3
  • RCM - EN 61326-1 Class A

BSP Inlet

  • 1/4” BSPT Female

BSP Outlet

  • 1/4” BSPT Female

Cable Length

  • 8.2 feet (2.5 metres) - Power Cable

Communication

  • Half-Duplex Bit Serial (Start-Stop Synchronization)
  • Modbus Protocol
  • RS-232C (9-Pins D-sub)
  • RS-485

Consumption

  • 70 VA, Max.

Display

  • 4-Digit LCD with CFL Backlight

Electrical Connectors

  • M3.5 Screw Terminals

Enclosure Finish

  • Off-White (Munsell 10Y7.5/0.5 or Equivalent) - Front Panel

Enclosure, Body Material

  • Steel - Blue, for Indoor Use

Flow Rate

  • 0.26 gpm (1 LPM) - for Purge Gas
  • 1 ± 0.5 LPM - for Sample Gas

Frequency

  • 50/60 Hz

Humidity

  • 90 % Relative Humidity, Non-Condensing, Max.

Input Signal

  • 0-1 Vdc from an O₂ Sensor Input Section is Non-Isolated - Analog Input Signal

Interface Connection

  • RS-485

Linearity

  • ± 1 % of Full Scale

Load

  • 100 kiloohms Min. for 0-1 Vdc - Permissible Load Resistance
  • 550 ohms, Max.

Materials of Construction

  • 304 Stainless Steel - Gas Inlet/Outlet, Sample Cell
  • Neoprene Rubber - Sample Cell
  • Toalon Tube - Internal Tubing

Measuring Range

  • 0 to 1,000 ppm Min., 0 to 100 Vol %, Max.
  • 0 to 200 ppm Min., 0 to 100 Vol %, Max.
  • 0 to 5 Vol % Min., 0 to 100 Vol %, Max.
  • 0 to 5 Vol % Min., 0 to 25 Vol %, Max.
  • 0 to 500 ppm Min., 0 to 100 Vol %, Max.
  • 0 to 500 ppm Min., 0 to 5,000 ppm, Max.
  • Dust: 100 µg/Nm³ or Less in Particle Size of 0.3 μm or Less

Media

  • Dry Air
  • Gas

Mounting

  • 19” Rack Mounting with Slide Rail

NPT Inlet

  • 1/4” NPT Female - Gas Inlet

NPT Outlet

  • 1/4” NPT Female - Gas Outlet

Operating Temperature

  • -5° to 45° C (23° to 113° F)

Output

  • 0-1 Vdc, Non-Isolated - Analog Output Signal
  • 4-20 mA DC, Non-Isolated - Analog Output Signal

Power Supply

  • 100-240 Vac - Voltage Rating
  • 85-264 Vac - Allowable Range

Pressure Range

  • psi: 0 to 1.45 (kPa: 0 to 10, bar: 0 to 0.1) or Less (Gas Outlet Side Should be Open to the Atmospheric Air)

Ratio

  • ≤ 1:20 (Built-in Paramagnetic O₂ Analyzer) - Measuring Range Ratio
  • ≤ 1:5 (Except Built-in Paramagnetic O₂ Analyzer) - Measuring Range Ratio

Relay Output

  • 250 Vac / 2 A, Resistive Load - Relay Contact Output

Repeatability

  • ± 0.5 % of Full Scale

Response Time

  • Within 15 seconds Including Replacement Time of Sample Gas (For 90 % of Full Scale Response) - for 1 or 2 Component Measurement
  • Within 30 seconds Including Replacement Time of Sample Gas (For 90 % of Full Scale Response) - for more than 2 Component Measurement

Span & Zero Effect

  • ± 2 % of Full Scale Per Week

Temperature Range

  • 0° to 50° C (32° to 122° F)

Window Material

  • CaF₂ - Infrared Ray Transmitting Window